IviScope Acquisition Start Time
IviScope Base Property
Data Type | Access | Applies to | Coercion | High Level VIs |
---|---|---|---|---|
R/W | N/A | None | IviScope Configure Acquisition Record |
Description
This properties specifies the length of time from the trigger event to the first point in the waveform record. The units are seconds. If the value is positive, the first point in the waveform record occurs after the trigger event. If the value is negative, the first point in the waveform record occurs before the trigger event.
Property Node Path
IviScope»Acquisition»Acquisition Start Time