IviScope TV Trigger Line Number [TV]
IviScopeTVTrigger Property [TV]
Data Type | Access | Applies to | Coercion | High Level VIs |
R/W | N/A | None | IviScope Configure TV Trigger Source |
Description
Specifies the line on which the oscilloscope triggers. This property affects instrument operation only when the IviScope Trigger Type property is set to IVISCOPE_VAL_TV_TRIGGER and when the IviScope TV Trigger Event [TV] property is set to IVISCOPE_VAL_LINE_NUMBER. The line number setting is independent of the field. This means that to trigger on the first line of the second field, you must set this property to the value of 263 (assuming that field one had 262 lines).
Property Node Path
IviScope»Trigger»TV Triggering [TV]»TV Trigger Line Number [TV]