IviScope Horizontal Record Length

LabVIEW IVI Class Driver

IviScope Horizontal Record Length

IviScope Base Property

Data Type Access Applies to Coercion High Level VIs
RO N/A N/A IviScope Actual Record Length

Description

Returns the actual number of points the oscilloscope acquires for each channel. The value is equal to or greater than the minimum number of points you specify IviScope Horizontal Minimum Number of Points property.

Allocate a array of this size or greater to pass as the Waveform Array parameter of the IviScope Read Waveform and IviScope Fetch Waveform VIs.

Note  Oscilloscopes may use different size records depending on the value specified for the IviScope Acquisition Type property.

Property Node Path

IviScope»Acquisition»Horizontal Record Length