IviDCPwr Triggered Current Limit [TRG]

LabVIEW IVI Class Driver

IviDCPwr Triggered Current Limit [TRG]

IviDCPwrTrigger Property [TRG]

Data Type Access Applies to Coercion High Level VIs
R/W Channel N/A IviDCPwr Configure Triggered Current Limit [TRG]

Description

This channel-based property specifies the value to which the power supply sets the current limit after a trigger event occurs. The units are amps.

After you call IviDCPwr Initiate [TRG], the power supply waits for a trigger event from the source you specify with the IviDCPwr Trigger Source [TRG], property. After a trigger event occurs, the power supply sets the current limit to the value of this property. After a trigger occurs, the value of the IviDCPwr Current Limit [TRG] property reflects the new value to which the current limit has been set.

Property Node Path

IviDCPwr»Trigger»Triggered Current [TRG]