IviDmm Sample Trigger [MP]

LabVIEW IVI Class Driver

IviDmm Sample Trigger [MP]

Multipoint Property

Data Type Access Applies to Coercion High Level VIs
R/W N/A None IviDmm Configure Multipoint

Description

Specifies the sample trigger source. This property affects instrument operation only when the IviDmm Sample Count [MP] property is greater than 1.

When the DMM takes a measurement and the IviDmm Sample Count [MP] property is greater than 1, the DMM does not take the next measurement until the event you specify with this property occurs.

When you set this property to IVIDMM_VAL_IMMEDIATE, the DMM does not wait for a trigger of any kind between measurements. When you set this property to IVIDMM_VAL_EXTERNAL, the DMM waits for a trigger on the external trigger input before it takes the next measurement. When you set this property to IVIDMM_VAL_SOFTWARE_TRIG, the DMM waits until you call the IviDmm Send Software Trigger [SWT] VI before it takes the next measurement. When you set this property to IVIDMM_VAL_INTERVAL, the DMM waits the length of time you specify with the IviDmm Sample Interval [MP] property before it takes the next measurement.

Property Node Path

IviDmm»Multipoint Acquisition»Sample Trigger [MP]

Defined Values

IVIDMM_VAL_IMMEDIATE

IVIDMM_VAL_EXTERNAL

IVIDMM_VAL_SOFTWARE_TRIG

IVIDMM_VAL_INTERVAL