Configuring AxoPatch 200 Patch Clamps

WinWCP

Getting Started > Amplifiers > Configuring AxoPatch 200 Patch Clamps

 

The Axon (now Molecular Devices) Axopatch 200A&B patch clamps provide gain and current-/voltage-clamp operating mode information via gain and mode telegraph outputs on the rear panel of the amplifier. Up to 4 Axopatch 200 patch clamps can be supported.

 

For each Axopatch 200 in use, connections must be made between the Scaled Output, 10 Vm and I Output signal output channels, Gain and Mode telegraph channels, and the analogue inputs of the laboratory interface. A connection must also be established between an analogue output and the Axopatch 200 Command Input.

 

In voltage-clamp mode, the Scaled Output channel monitors the current signal and the 10 Vm provides the cell membrane potential. In current-clamp mode, the Scaled Output switches to membrane potential and the I Output output is monitored for the current. The current- voltage-clamp mode switching between the 10 Vm and I Output outputs can be accomplished either by physically changing the signal cable attachments or by reassigning analogue inputs using software.

 

Physically switched Secondary Input Switching

 

To configure WinWCP to use an Axopatch 200 as Amplifier #1, select Axopatch 200  as the amplifier type on the Amplifier #1 page of the Input Channels & Amplifiers Setup dialog box, as shown below.

 

 

The default connections are:

 

 

Primary Ch.

Secondary Ch.

 

 

 

Axopatch200

Scaled Output

10 Vm

I Output

Mode Telegraph

Gain Telegraph

Command Input

Laboratory
Interface

AI 0

AI 1

AI 1

AI 6

AI 7

AO 0

 

The secondary input channel (AI1) must be physically switched between 10 Vm and Im when the patch clamp is switched from voltage- to current-clamp mode. This can be accomplished by physically swapping the coaxial signal cables or using a BNC switch box.

 

Software switched Secondary Channel Switching

 

Voltage/current-clamp mode switching of the 10 Vm and I Output channels can be accomplished automatically by attaching the 10 Vm and I Output outputs to different analogue input channels and remapping the analogue input to the secondary channel (Ch.1) using software.

 

Software channel switching can be configured by selecting the IClamp option on the Amplifier #1 page and selecting a free analogue input (e.g. AI 4) for the Ch.1 Secondary Channel, as shown below.

 

 

The connections for a software switched secondary channel are shown below.

 

 

Primary Ch.

Secondary Ch.

 

 

 

Axopatch
200

Scaled Output

10 Vm

I Output

Mode Telegraph

Gain Telegraph

Command Input

Laboratory
Interface

AI 0

AI 1

AI 4

AI 6

AI 7

AO 0

 

Multiple Axopatch 200 amplifiers.

 

Up to three additional Axopatch 200 amplifiers can be defined as Amplifiers #1-#3.

 

The default connections for Amplifier #2 in a two amplifier configuration are shown below

 

 

Primary Ch.

Secondary Ch.

 

 

 

Axopatch
200

Scaled Output

10 Vm

I Output

Mode Telegraph

Gain Telegraph

Command Input

Laboratory
Interface

AI 2

AI 3

AI 3

AI 4

AI 5

AO 1

 

This is a physically switched secondary channel configuration and assumes that Amplifier #1  is the same. To select a software switched configuration, the gain and mode telegraph channels must be moved to input channels 7-15 (if the laboratory interface supports them) or disabled by setting the Gain or Mode Telegraph settings to Off, freeing up the input channels for use as current-clamp secondary inputs.