IVISCOPE_ATTR_VERTICAL_RANGE

CVI/LabWindows IVI Class Driver

IVISCOPE_ATTR_VERTICAL_RANGE

IviScopeBase Attribute

Data Type Access Applies to Coercion High Level Functions Id Definition
ViReal64 R/W Channel Up IviScope_ConfigureChannel 1250001

Description

Specifies the absolute value of the input range for a channel. The units are volts.

For example, to acquire a sine wave which spans between -5.0 and 5.0 V, set the IVISCOPE_ATTR_VERTICAL_RANGE attribute to 10.0 V.