IVIPWRMETER_ATTR_INTERNAL_TRIGGER_LEVEL

CVI/LabWindows IVI Class Driver

IVIPWRMETER_ATTR_INTERNAL_TRIGGER_LEVEL

IviPwrMeterInternalTrigger Attribute

Data
type
Access Applies to Coercion High Level Functions
ViReal64 R/W N/A None IviPwrMeter_ConfigureInternalTriggerLevel

Description

Specifies the trigger level for the measurement signal. The power meter leaves the Idle state when the measurement signal on the channel specified by the IVIPWRMETER_ATTR_INTERNAL_TRIGGER_EVENT_SOURCE [IT] attribute crosses the value specified by this attribute. The value of this attribute is specified in the same unit as the value of the IVIPWRMETER_ATTR_UNITS.

This attribute affects the behavior of the instrument only if the IVIPWRMETER_ATTR_TRIGGER_SOURCE [TRG] attribute is set to Internal.