Analyzer - Using - Settings - Display

DeviceNet Measurement & Automation Explorer

Configure Display Settings

Complete the following steps to configure the display settings:

  1. If the analyzer is capturing, stop capture. Refer to Start and Stop Capture for details.
  2. Select Settings»Display Settings... to open the display settings dialog.
  3. Configure the display settings, including:
    • Update Rate—A rate at which the UI updates (in milliseconds).
    • You can select a value from an enumerated list, including 1000 ms, 2000 ms, and 5000 ms, where 1000 ms is the default value.

    • Time Stamp Format—Time stamp of captured packets listed in the Packet List window, either absolute or relative format.
      • Absolute—In local system time format. An example is shown below.
      • Format 14:27:13.257.3
        Value 14 27 13 257 3
        Description Hour Minute Second Millisecond 100*Microsecond
        This packet is captured in 14:27:13.257.3 (local time zone).
      • Relative—In seconds since the first packet captured. An example is shown below.
      • Format(Unit : Second) 285.2345
        Description This packet is captured 285.2345 seconds after the first packet was captured.
  4. Click OK.