EDVR Error with Ibcnt Set to 0xE1060078 (-519700360)

NI-488.2

EDVR Error with Ibcnt Set to 0xE1060078 (-519700360)

This error occurs if you are using a GPIB-ENET/100 and the network link is broken between the host and the GPIB-ENET/100 interface. This error can be returned for one of the following reasons:

  • The cables are broken or removed.
  • The GPIB-ENET/100 is switched off during communication.
  • The virtual connection between your host computer and the GPIB-ENET/100 has encountered unrecoverable errors.

Return to Troubleshooting EDVR Error Conditions