RF Signal Analyzer Calibration

NI PXI-5660

Calibration

Every NIĀ 5660 RF Vector Signal Analyzer is individually calibrated for accurate frequency response at the factory and ships with a calibration certificate verifying NIST-traceable accuracy levels.

During frequency-response calibration, the RF Signal Analyzer is used to measure a NIST-certified high-precision signal. Any error in the returned data is quantified as a set of calibration constants. These calibration constants are used by the software to calculate and apply correction to your analysis based upon the spectrum of interest. For more information on applying calibration correction, refer to the example programs installed with the RF Signal Analyzer.

To preserve specified accuracy and NIST traceability, NI recommends returning both modules of the RF Signal Analyzer to the factory for annual recalibration. The RF downconverter module and the IF digitizer module are calibrated independently of one another, not as a combined system. At this time, recalibration of both modules can only be performed by NI. For more information on calibration, contact NI or visit ni.com/calibration.