DAQmxAdjustDSAAICal
int32 DAQmxAdjustDSAAICal (uInt32 calHandle, float64 referenceVoltage);
Purpose
Adjusts the external calibration constants for the analog input section of a DSA device. You must connect a known voltage to the device and specify that voltage with the referenceVoltage parameter.
Parameters
Input | ||
Name | Type | Description |
---|---|---|
calHandle | uInt32 | A reference to the calibration session that you created using the DAQmxInitExtCal function. |
referenceVoltage | float64 | The known voltage, in volts, to use as a reference for calibration. This voltage should be between +4.9 V and +9.1 V. |
Return Value
Name | Type | Description |
---|---|---|
status | int32 | The error code returned by the function in the event of an error or warning. A value of 0 indicates success. A positive value indicates a warning. A negative value indicates an error. |