Input Process Data in Detail

BECKHOFF KL336x

 
BECKHOFF KL3361, KL3362: KS2000 Configuration Software

Input Process Data in Detail

Data word 0

Here you can specify which input value is transferred with data word 0 from the oscilloscope terminal to the control.

Input value Comment
Current measurement reading Current analog value
Maximum value Maximum value of a recording
Minimum value Minimum value of a recording
RMS value RMS value of a recording: Sqrt((Sum(xn2))/n
Mean value Arithmetic mean of a recording: (Sum (xn))/n
Peak to peak value Peak to peak value of a recording
Cycle duration Cycle duration of a recording
Pulse width HIGH
  • four successive values have to be above the threshold level for activating the start
  • four successive values have to be below the threshold level for activating the stop
Pulse width LOW
  • four successive values have to be above the threshold level for activating the start
  • four successive values have to be below the threshold level for activating the stop
Mark/space ratio Duty cycle
Jitter Tmax reserved
Jitter Tmin reserved
Jitter Tmean reserved
Histogram Max Most frequent value of a recording
Latched value timer 1 stored value of the timer (chronometer) from trigger unit 1* *) for KL3362 also trigger unit 3 (via data word 2 or 3)
Online value timer 1 current value of the timer (chronometer) from trigger unit 1*
Latched value timer 2 stored value of the timer (chronometer) from trigger unit 2** **) for KL3362 also trigger unit 4 (via data word 2 or 3)
Online value timer 2 current value of the timer (chronometer) from trigger unit 2**
Error counter inner envelope curve error counter of the inner envelope curve
Error counter outer envelope curve error counter of the outer envelope curve
Number of samples up to analog value greater than threshold level Number of measurement points recorded up to the time when the analog value exceeded the threshold level.
Number of samples up to analog value less than threshold level Number of measurement points recorded up to the time when the analog value was less than the threshold level.
Data word 1

Here you can specify which input value is transferred with data word 1 from the oscilloscope terminal to the control.

(input value see Data word 0).

Data word 2 (KL3362 only)

Here you can specify which input value is transferred with data word 2 from the oscilloscope terminal to the control.

(input value see Data word 0).

Data word 3 (KL3362 only)

Here you can specify which input value is transferred with data word 3 from the oscilloscope terminal to the control.

(input value see Data word 0).