External Calibration
Owning Palette: CalibrationUse the NI-DMM External Calibration VIs to perform an external calibration on the DMM.
Palette Object | Description |
---|---|
niDMM Initialize External Cal |
The following operations are performed if the Calibration Password is valid: - Creates a new session for external calibration to the device you specify for the Instrument Descriptor parameter. - Resets the device and prepares the EEPROM for new calibration coefficients. - Returns a ViSession handle that you use to identify the instrument in all calibration adjustments and post-adjustment verification steps. |
niDMM Close External Cal |
Performs the specified Action, closes the specified external calibration session obtained from niDMM Initialize External Cal, and deallocates resources that it reserved. |
niDMM Cal Adjust Gain |
Calibrates the gain coefficient for the supplied Function, Range, and Input Resistance. |
niDMM Cal Adjust Offset |
Calibrates the offset and Auto Zero offset for the supplied Function, Range, and Input Resistance. |
niDMM Cal Adjust AC Filter |
(NI 4070/4071/4072 only) Calibrates the filter coefficients used for AC measurements of the supplied Function and Range. |
niDMM Cal Adjust Linearization |
(NI 4065 only) Compensates for any non-linearities. |
niDMM Cal Adjust Miscellaneous |
Performs a specialized calibration step depending on the specified Type. |
niDMM Cal Adjust LC |
For the NI 4072 only, performs a specialized LC calibration step depending on the specified Stage. |