Calibration Utility
Owning Palette: CalibrationUse the NI-DMM Calibration Utility VIs for optional functionality when performing a calibration.
Palette Object | Description |
---|---|
niDMM Get Cal Count |
Returns the calibration count for the specified type of calibration. |
niDMM Get Cal Date and Time |
Returns the date and time of the last calibration performed. |
niDMM Get Last Cal Temp |
Returns the temperature during the last calibration procedure. |
niDMM Get Dev Temp |
Returns the current temperature of the device. |
niDMM Get External Cal Recommended Interval |
Returns the recommended interval between external recalibration in months. |
niDMM Get Self Cal Supported |
Returns a Boolean value that expresses whether the DMM that you are using can perform self-calibration. |
niDMM Get Cal User Defined Info Max Size |
Returns the maximum string length that can be stored in the EEPROM. |
niDMM Get Cal User Defined Info |
Returns the user-defined calibration information stored in the EEPROM. |
niDMM Restore Last External Cal Constants |
Reverts the device to the calibration constants from the last complete external calibration. This VI recovers the hardware if a fatal system error occurs during an external or self-calibration procedure. |
niDMM Set Cal Password |
Changes the password required to enable external calibration functionality. The maximum password string length is eight characters, excluding the termination character. "NI" is the default password. A password is required for external calibration. Be sure to record the password in a secure location. |
niDMM Set Cal User Defined Info |
Stores the user-defined information in the EEPROM. |