Voltage Limit Range

NI DC Power Properties

DC Current:Voltage Limit Range

Short Name: Voltage Limit Range

Specifies the voltage limit range for the specified channel(s). The range defines the valid values to which the voltage limit can be set. Use the Voltage Limit Autorange property to enable automatic selection of the voltage limit range.

The Voltage Limit Range property is applicable only if the Output Function property is set to DC Current.

The channel must be enabled for the specified voltage limit range to take effect. Refer to the Output Enabled property for more information about enabling the output channel.

Valid Values:

NI PXI-4110
ChannelVoltage Limit Range (V)Voltage Limit (V)
060 to +6
1, 2200 to +20
2200 to +20

NI PXI-4130
ChannelVoltage Limit Range (V)Voltage Limit (V)
060 to +6
160 to +6
200 to +20
Note Note   If a range other than what is listed in the preceding table is selected, it will be coerced to the next-highest range. For example, requesting the 10 V voltage limit range on Channel 1 on the NI-PXI 4130 coerces the voltage limit range to 20 V. Refer to the Ranges topic in the NI DC Power Supplies and SMUs Help for more information about coercion.

Remarks

The following table lists the characteristics of this property.

Data Type ViReal64
Permissions R/W
Channel Based Yes
High-Level VI niDCPower Configure Voltage Limit Range