Voltage Limit Autorange

NI DC Power Properties

DC Current:Voltage Limit Autorange

Short Name: Voltage Limit Autorange

Specifies whether or not NI-DCPower automatically selects the voltage limit range based on the desired voltage limit for the specified channel(s).

If this property is set to On, NI-DCPower ignores any changes you make to the Voltage Limit Range property. If you change the Voltage Limit Autorange property from On to Off, NI-DCPower remembers the last value that the Voltage Limit Range property was set to (or the default value if it was never set) and uses that value as the voltage limit range.

Read the Voltage Limit Range property to find out which range NI-DCPower automatically selects.

The Voltage Limit Autorange property is applicable only if the Output Function property is set to DC Current.

Defined Values:

OnNI-DCPower automatically selects the voltage limit range.
OffNI-DCPower does not automatically select the voltage limit range.

Default Value: Off

Remarks

The following table lists the characteristics of this property.

Data Type ViInt32
Permissions R/W
Channel Based Yes
High-Level VI None