DAQmx DSA Calibration VIs

DAQmx - Data Acquisition VIs and Functions

DAQmx DSA Calibration VIs

Owning Palette: DAQmx Calibration VIs and Functions

Use the DAQmx DSA Calibration VIs to perform external calibrations of the analog input section, analog output section, and timebase of DSA devices.

Palette Object Description
DAQmx Adjust DSA AI Calibration Adjusts the external calibration constants for the analog input section of a DSA device. You must connect a known voltage to the device and specify that voltage with reference voltage.
DAQmx Adjust DSA AO Calibration Adjusts the external calibration constants for the analog output section of a DSA device. You must use the device to generate a high voltage and low voltage at a specified gain, measure the high and low voltages, then specify the requested high voltage, low voltage, and gain along with the actual high voltage and low voltage.
DAQmx Adjust DSA Timebase Calibration Adjusts the external calibration constant for the timebase of a DSA device with an adjustable oscillator. You must connect a sinusoidal signal with a known frequency to the device and specify that frequency with reference frequency.