DAQmx Adjust SCXI Calibration

DAQmx - Data Acquisition VIs and Functions

DAQmx Adjust SCXI Calibration

Adjusts the external calibration constants for an SCXI module. The instances of this polymorphic VI correspond to the SCXI module you want to calibrate. Refer to the calibration procedure for your module for detailed calibration instructions.

Note  After you commit the calibration, you cannot undo it. If any calibration settings are incorrect, you must recalibrate the module.

Use the pull-down menu to select an instance of this VI.

minibutton.gif  Place on the block diagram. minibutton.gif  Find on the Functions palette.

1102 

Adjusts the external calibration constants for an SCXI-1102 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 1102 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1104 

Adjusts the external calibration constants for an SCXI-1104 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 1104 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1112 

Adjusts the external calibration constants for an SCXI-1112 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 1112 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1122 

Adjusts the external calibration constants for an SCXI-1122 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 1122 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1124 

Adjusts the external calibration constants for an SCXI-1124 module. You must use the DAQmx Setup SCXI Calibration VI to generate a voltage or current. Measure that voltage or current, and specify the measured value with measured output.

DAQmx Adjust SCXI 1124 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif measured output specifies in volts or amperes the voltage or current measured at the output channel specified in the DAQmx Setup SCXI Calibration VI.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1125 

Adjusts the external calibration constants for an SCXI-1125 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 1125 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1126 

Adjusts the external calibration constants for an SCXI-1126 module. You must connect a sinusoidal signal with a known frequency to the device and specify that frequency with reference frequency.

DAQmx Adjust SCXI 1126 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference frequency specifies in hertz the frequency of the signal to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1141 

Adjusts the external calibration constants for an SCXI-1141 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 1141 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1142 

Adjusts the external calibration constants for an SCXI-1142 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 1142 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1143 

Adjusts the external calibration constants for an SCXI-1143 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 1143 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1502 

Adjusts the external calibration constants for an SCXI-1502 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 1502 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1503 

Adjusts the external calibration constants for an SCXI-1503 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 1503 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1503 Current 

Adjusts the current calibration constants for an SCXI-1503 module. You must measure the current generated on the physical channel you specify and specify the measured current in measured output.

DAQmx Adjust SCXI 1503 Current Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
civrn.gif physical channel specifies the physical channel to calibrate.
cdbl.gif measured current specifies in amperes the current measured at the specified physical channel.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1520 

Adjusts the external calibration constants for an SCXI-1520 module. This module requires reference signals of 0.0 V at gains of 1, 15, 20, and 310 on a particular channel to perform an offset calibration for that channel.

DAQmx Adjust SCXI 1520 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1521 

Adjusts the external calibration constants for an SCXI-1521 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 1521 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

153x 

Adjusts the external calibration constants for an SCXI-1530 or SCXI-1531 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 153x Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.

1540 

Adjusts the external calibration constants for an SCXI-1540 module. You must connect a known voltage to the module and specify that voltage with reference voltage.

DAQmx Adjust SCXI 1540 Calibrationc.gif

cu32.gif calhandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration VI.
cstr.gif input calibration source specifies the input source selection.
Loopback0 Loopback the internal excitation voltage with 0 degree phase shift.
Loopback180 Loopback the internal excitation voltage with 180 degree phase shift.
Ground Connect the channel to ground.
cdbl.gif reference voltage specifies in volts the known voltage to use as a reference for calibration.
cdbl.gif measured output specifies in volts the voltage measured at the output of the module.
cerrcodeclst.gif error in describes error conditions that occur before this VI or function runs. The default is no error. If an error occurred before this VI or function runs, the VI or function passes the error in value to error out. If an error occurs while this VI or function runs, the VI or function runs normally and sets its own error status in error out. Use the Simple Error Handler or General Error Handler VIs to display the description of the error code. Use error in and error out to check errors and to specify execution order by wiring error out from one node to error in of the next node.
cbool.gif status is TRUE (X) if an error occurred before this VI or function ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI or function ran. The default is FALSE.
ci32.gif code is the error or warning code. The default is 0. If status is TRUE, code is a negative error code. If status is FALSE, code is 0 or a warning code.
cstr.gif source identifies where an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.
iu32.gif calhandle out is a reference to the calibration session. Wire this output to other external calibration VIs.
ierrcodeclst.gif error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, error out describes the error status that this VI or function produces. Right-click the error out indicator on the front panel and select Explain Error from the shortcut menu for more information about the error.
ibool.gif status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.
ii32.gif code is the error or warning code. If status is TRUE, code is a nonzero error code. If status is FALSE, code is 0 or a warning code.
istr.gif source identifies where and why an error occurred. The source string includes the name of the VI that produced the error, what inputs are in error, and how to eliminate the error.