等待触发状态(OPC?)

E5071C

等待触发状态(OPC?)

有关样本程序的其他主题

概述

这个样本程序说明如何使用:TRIG:SING命令,直到完成测量周期。

  • 这个样本程序在通道/迹线的最大数量设置为9时能正确运行。

这个样本程序使用:TRIG:SING命令开始一个扫描(测量)周期,使用*OPC命令等待测量周期完成,然后打印消息并退出。

参见这个程序的等待测量结束

Excel VBA中的样本程序
 

Sub trg_sing_Click()

    Dim defrm As Long

    Dim vi As Long

    Dim ContMode(9) As String

    Dim Result As String * 10

    Dim i As Integer

    Const TimeOutTime = 100000 ' TimeOut time should be greater than the measurement time.

    '

    ' Assign a GPIB address to the I/O pass.

    Call viOpenDefaultRM(defrm)

    Call viOpen(defrm, "GPIB0::17::INSTR", 0, 0, vi)

    Call viSetAttribute(vi, VI_ATTR_TMO_VALUE, TimeOutTime)

    '

    ' Store the settings of continuous initiation mode for eachchannel

    ' (on for channels 1 and 2; off for channels 3 through 9)

    ' into the array variable ContMode().

    ContMode(1) = "ON"

    ContMode(2) = "ON"

    For i = 3 To 9

        ContMode(i) = "OFF"

    Next i

    '

    ' Turn on or off continuous initiation mode for each channel

    ' depending on the value of ContMode(*).

    For i = 1 To 9

        Call viVPrintf(vi, ":INIT" & CStr(i) & ":CONT " & ContMode(i) & vbLf, 0)

    Next i

    ' Set the trigger source to Bus Trigger.

    Call viVPrintf(vi, ":TRIG:SOUR BUS" & vbLf, 0)

    '

    ' Trigger the instrument to start a sweep cycle.

    Call viVPrintf(vi, ":TRIG:SING" & vbLf, 0)

    '

    ' Execute the *OPC? command and wait until the command

    ' returns 1 (i.e., the measurement cycle is completed).

    Call viVPrintf(vi, "*OPC?" & vbLf, 0)

    Call viVScanf(vi, "%t", Result)

    '

    ' Display a measurement completion message.

    Stat = MsgBox("Measurement complete", vbOKOnly)

    Call viClose(vi)

    Call viClose(defrm)

End Sub

HT Basic中的样本程序(trg_sing.htb)

10 DIM Cont_mode$(1:9)[9],Buff$[9]

20 INTEGER I

30 !

40 ASSIGN @Agte507x TO 717

50 !

60 Cont_mode$(1)="ON"

70 Cont_mode$(2)="ON"

80 Cont_mode$(3)="OFF"

90 Cont_mode$(4)="OFF"

100 Cont_mode$(5)="OFF"

110 Cont_mode$(6)="OFF"

120 Cont_mode$(7)="OFF"

130 Cont_mode$(8)="OFF"

140 Cont_mode$(9)="OFF"

150 !

160 FOR I=1 TO 9

170 OUTPUT @Agte507x;":INIT"&VAL$(I)&":CONT "&Cont_mode$(I)

180 NEXT I

190 OUTPUT @Agte507x;":TRIG:SOUR BUS"

200 !

210 OUTPUT @Agte507x;":TRIG:SING"

220 OUTPUT @Agte507x;"*OPC?"

230 ENTER @Agte507x;Buff$

240 !

250 PRINT "Measurement complete"

260 END