niFgen_ConfigureArbSequence

NI-FGEN C Function

niFgen_ConfigureArbSequence

ViStatus niFgen_ConfigureArbSequence (ViSession vi, ViInt32 sequenceLength, ViInt32[] waveformHandlesArray, ViInt32[] loopCountsArray, ViInt32* sequenceHandle);

Purpose

Configures the signal generator attributes that affect arbitrary sequence generation. These attributes are the NIFGEN_ATTR_ARB_SEQUENCE_HANDLE, NIFGEN_ATTR_ARB_GAIN, and NIFGEN_ATTR_ARB_OFFSET

Notes  The signal generator must not be in the Generating state when you call this function.
You must call the niFgen_ConfigureOutputMode function to set the output mode to NIFGEN_VAL_OUTPUT_SEQ before calling this function.

Parameters

Input
Name Type Description
vi ViSession Identifies your instrument session. vi is obtained from the niFgen_init or the niFgen_InitWithOptions functions and identifies a particular instrument session.
channelName ViConstString Specifies the channel name from which you want to configure an arbitrary sequence.

Default Value: "0"

sequenceHandle ViInt32 Specifies the handle of the arbitrary sequence that you want the signal generator to produce. NI-FGEN sets the NIFGEN_ATTR_ARB_SEQUENCE_HANDLE attribute to this value. You can create an arbitrary sequence using the niFgen_CreateArbSequence function. The niFgen_CreateArbSequence function returns a handle that you use to identify the sequence.

Default Value: None

gain ViReal64 Specifies the factor by which the signal generator scales the arbitrary waveform data. When you create arbitrary waveforms, you must first normalize the data points to a range of –1.00 to +1.00 V. You can use this parameter to scale the arbitrary waveform to other ranges. NI-FGEN sets the NIFGEN_ATTR_ARB_GAIN attribute to this value.

For example, to configure the output signal to range from –2.00 to +2.00 V, set the gain to 2.00.

Units: volts

Default Value: None

offset ViReal64 Specifies the value the signal generator adds to the arbitrary waveform data. When you create arbitrary waveforms, you must first normalize the data points to a range of –1.00 to +1.00 V. You can use this parameter to shift the range of the arbitrary waveform. NI-FGEN sets the NIFGEN_ATTR_ARB_OFFSET attribute to this value.

For example, to configure the output signal to range from 0.00 to 2.00 V instead of –1.00 to 1.00 V, set the offset to 1.00.

Units: volts

Default Value: None

Return Value

Name Type Description
Status ViStatus Returns the status code of this operation. The status code either indicates success or describes an error or warning condition. You can examine the status code from each call to an NI-FGEN function to determine if an error occurred. To obtain a text description of the status code, call the niFgen_error_message function. To obtain additional information about the error condition, call the niFgen_GetError function. To clear the error information from NI-FGEN, call the niFgen_ClearError function.

The general meaning of the status code is as follows:

Value Meaning
0 Success
Positive Values Warnings
Negative Values Errors