niDMM_CalAdjustMisc
ViStatus = niDMM_CalAdjustMisc(ViSession Instrument_Handle, ViInt32 Type)
Purpose
Performs a specialized calibration step depending on the specified Type.
 |
Note The NI 4050 and NI 4060 are not supported.
Refer to the NI 4065 6½ Digit DMM Calibration Procedure, the NI 4070/4072 6½ Digit FlexDMM Calibration Procedure, or the NI 4071 7½–Digit FlexDMM Calibration Procedure before using this function. |
Parameters
Input |
Name |
Type |
Description |
Instrument_Handle |
ViSession |
The session handle that you obtain from niDMM_InitExtCal. The handle identifies a particular instrument calibration session.
|
Type |
ViInt32 |
Specifies which of the miscellaneous calibration steps to perform.
NIDMM_EXTCAL_MISCCAL_VREF (default) |
Calibrate the Voltage Reference. |
NIDMM_EXTCAL_MISCCAL_RREF |
Calibrate the Resistance Reference. |
NIDMM_EXTCAL_MISCCAL_ZINT |
Calibrate the Internal Impedance. |
NIDMM_EXTCAL_MISCCAL_2WIRELEAKAGE |
Calibrate the 2-wire Leakage resistance. |
NIDMM_EXTCAL_MISCCAL_4WIRELEAKAGE |
Calibrate the 4-wire Leakage resistance. |
NIDMM_EXTCAL_MISCCAL_SECTION |
Update calibration information and verify calibration completeness. |
|
Return Value
Name |
Type |
Description |
Status |
ViStatus |
Reports the Status of this operation. To obtain a text description of the status code, call
niDMM_error_message. To obtain additional information concerning the error condition, use
niDMM_GetError. |