Holdoff






Public Data Item |
Declared in: NiIviScope.h |
Declaration
double Holdoff;
Description
Specifies the length of time in seconds that the oscilloscope waits after it detects a trigger until it enables the trigger subsystem to detect another trigger.
This property affects instrument operation only when the oscilloscope requires multiple acquisitions to build a complete waveform. The oscilloscope requires multiple waveform acquisitions when it uses equivalent-time sampling or when you set the Acquisition.Type property to CNiIviScope::Envelope or CNiIviScope::Average.