CNiIviScopeReferenceLevel::Mid

Measurement Studio IVI

Class
CNiIviScopeReferenceLevel::
Mid
Base ClassesData ItemsFunctionsGo to hierarchy chart    Prev page: LowNext page: Configure    
Public Data Item Declared in:
NiIviScope.h

'Declaration' icon -- Shortcut to top of page. Declaration

double Mid;

'Description' icon -- Shortcut to top of page. Description

Specifies the mid reference the oscilloscope uses for waveform measurements. The value is a percentage of the difference between the Voltage High and Voltage Low. The oscilloscope calculates the Voltage High and the Voltage Low using either the min/max or histogram methods. The min/max method uses the maximum and minimum values found. The histogram method uses the most common values found above and below the middle of the waveform.

'See Also' icon -- Shortcut to top of page. See Also