[SENSe:]TOTalize:DATA?

From Keysight 34970A/34972A

[SENSe:]TOTalize:DATA?

Syntax

[SENSe:]TOTalize:DATA? [(@<ch_list>)]

Description

This query reads the count on the specified totalizer channels. If you have configured the count to be reset when it is read (see CONFigure:TOTalize and [SENSe:]TOTalize:TYPE commands), then this command will reset the count to 0 after it is read. The count is reset regardless of whether the specified channels are in a scan list or even whether a scan is in progress.

Used With:

  • 34907A   Multifunction Module (totalizer channel only)

Parameters

Name Type Range of Values Default Value
<ch_list> Channel List

One or more totalizer channels, as shown:

(@303) - channel 03 on the module in slot 300.

(@103,203,303) - channel 03 on the modules on slot 100, 200, and 300.

If you omit the optional <ch_list> parameter, this command applies to the currently defined scan list.

Remarks

  • The maximum count is 67,108,863 (226 - 1). The count rolls over to 0 after reaching the maximum allowed value.
  • The output from this command is affected by the settings of the FORMat:READing commands. Depending on the formats selected, each reading may or may not be stored with measurement units, time stamp, channel number, and alarm status information.

Return Format

The query returns an unsigned decimal value representing the count on each totalizer channel specified (a full 32-bit count is returned). Multiple responses are separated by commas.

Example

The following query reads the count on totalizer channel 03 on the module in slot 3.

TOT:DATA? (@303)

Typical Response: 1.32130000E+03

 

See Also

SENSe Subsystem Introduction

CONFigure:TOTalize

[SENSe:]TOTalize:TYPE