[SENSe:]TOTalize:SLOPe

From Keysight 34970A/34972A

[SENSe:]TOTalize:SLOPe

Syntax

[SENSe:]TOTalize:SLOPe <edge>[,(@<ch_list>)]

[SENSe:]TOTalize:SLOPe? [(@<ch_list>)]

Description

This command configures the specified totalizer channels to count on the rising edge (positive) or falling edge (negative) of the input signal.

Used With:

  • 34907A   Multifunction Module (totalizer channel only)

Parameters

Name Type Range of Values Default Value
<edge> Discrete {NEGative|POSitive} This is a required parameter.
<ch_list> Channel List

One or more totalizer channels, as shown:

(@303) - channel 03 on the module in slot 300.

(@103,203,303) - channel 03 on the modules on slot 100, 200, and 300.

If you omit the optional <ch_list> parameter, this command applies to the currently defined scan list.

Remarks

  • The selected slope is stored in volatile memory and will be set to "POS" when power is turned off or after a Factory Reset (*RST command).

Return Format

The query returns NEG or POS for the specified channels. Multiple responses are separated by commas .

Examples

The following command configures totalizer channel 03 on the module in slot 300 to count on the negative edge (falling) of the input signal.

TOT:SLOP NEG,(@303)

 

The following query returns the edge setting on totalizer channel 3 on the modules in slots 1 and 3.

TOT:SLOP? (@103,303)

Typical Response: NEG,NEG

See Also

SENSe Subsystem Introduction