ROUTe:MONitor:DATA?

From Keysight 34970A/34972A

ROUTe:MONitor:DATA?

Syntax

ROUTe:MONitor:DATA?

Description

This query reads the monitor data from the selected channel. It returns the reading only; the units, time, channel, and alarm information are not returned (the FORMat:READing commands do not apply to monitor readings).

Remarks

  • If the Monitor mode is not currently enabled, this query returns an error indicating that it is unable to perform the requested operation.
  • Readings acquired during a Monitor are not stored in memory but they are displayed on the front panel; however, all readings from a scan in progress at the same time are stored in memory.

Return Format

The query returns one reading for the Monitor channel, in the format +1.12345678E+01. If no data is available for the specified channel, instrument waits for data to become available.

Examples

The following query returns one reading from the Monitor channel.

ROUT:MON:DATA?

Typical Response: +1.84280000E-05

See Also

ROUTe Subsystem Introduction

ROUTe:MONitor

ROUTe:MONitor:STATe