OUTPut:ALARm{1|2|3|4}:CLEar
OUTPut:ALARm{1|2|3|4}:CLEar
OUTPut:ALARm:CLEar:ALL
Syntax
OUTPut:ALARm{1|2|3|4}:CLEar
OUTPut:ALARm:CLEar:ALL
Description
The OUTPut:ALARm{1|2|3|4}:CLEar command clears the specified alarm output line. The OUTPut:ALARm:CLEar:ALL command clears all four alarm output lines.
Used With:
- 34901A 20 Channel Multiplexer (2/4-wire) Module
- 34902A 16 Channel Multiplexer (2/4-wire) Module
- 34907A Multifunction Module (digital input only)
- 34908A 40 Channel Single-Ended Multiplexer Module
Remarks
-
You can manually clear the output lines at any time (even during a scan) and the alarm data in memory is not cleared (however, data is cleared when you initiate a new scan). The alarm outputs are also cleared when you initiate a new scan.
-
A Factory Reset (*RST command) clears all four alarm outputs but does not clear the alarm queue in either configuration.
Examples
The following command clears alarm output line 2.
OUTP:ALAR2:CLE
The following command clears all four alarm output lines.
OUTP:ALAR:CLE:ALL