OUTPut:ALARm{1|2|3|4}:CLEar

From Keysight 34970A/34972A

OUTPut:ALARm{1|2|3|4}:CLEar
OUTPut:ALARm:CLEar:ALL

Syntax

OUTPut:ALARm{1|2|3|4}:CLEar

OUTPut:ALARm:CLEar:ALL

Description

The OUTPut:ALARm{1|2|3|4}:CLEar command clears the specified alarm output line. The OUTPut:ALARm:CLEar:ALL command clears all four alarm output lines.

Used With:

  • 34901A   20 Channel Multiplexer (2/4-wire) Module
  • 34902A   16 Channel Multiplexer (2/4-wire) Module
  • 34907A   Multifunction Module (digital input only)
  • 34908A   40 Channel Single-Ended Multiplexer Module

Remarks

  • You can manually clear the output lines at any time (even during a scan) and the alarm data in memory is not cleared (however, data is cleared when you initiate a new scan). The alarm outputs are also cleared when you initiate a new scan.

  • A Factory Reset (*RST command) clears all four alarm outputs but does not clear the alarm queue in either configuration.

Examples

The following command clears alarm output line 2.

OUTP:ALAR2:CLE

 

The following command clears all four alarm output lines.

OUTP:ALAR:CLE:ALL

See Also

OUTPut Subsystem Introduction

OUTPut:ALARm:MODE

OUTPut:ALARm:SLOPe

OUTPut:ALARm{1|2|3|4}:SOURce