Save Homogeneity Calculations

EPMA Xtreme Probe

 

Probe Menu Details > Output

Save Homogeneity Calculations

This menu (homogeneity output format) will create an output file of standard or unknown average weight, oxide and atomic percents along with homogeneity calculations and also the average detection and analytical sensitivity for each data point.

 

Please note that the HTO (homogeneity test of) is calculated yes, no or null at the 1% concentration level.

 

SAMPLE      Sample number and name (in double quotes)

El AT%      Atomic percent

El WT%      Elemental weight percent

Ox WT%      Oxide weight percent

TOTAL       Total weight percent

 

El HRO60    Elemental t-test “homogenity range of” (60 percent confidence)

El HRO80    Elemental t-test “homogenity range of” (80 percent confidence)

El HRO90    Elemental t-test “homogenity range of” (90 percent confidence)

El HRO95    Elemental t-test “homogenity range of” (95 percent confidence)

El HRO99    Elemental t-test “homogenity range of” (99 percent confidence)

 

El HTO60    Elemental t-test “homogenity test of” (60 percent confidence)

El HTO80    Elemental t-test “homogenity test of” (80 percent confidence)

El HTO90    Elemental t-test “homogenity test of” (90 percent confidence)

El HTO95    Elemental t-test “homogenity test of” (95 percent confidence)

El HTO99    Elemental t-test “homogenity test of” (99 percent confidence)

 

El HLO60    Elemental t-test “homogenity level of” (60 percent confidence)

El HLO80    Elemental t-test “homogenity level of” (80 percent confidence)

El HLO90    Elemental t-test “homogenity level of” (90 percent confidence)

El HLO95    Elemental t-test “homogenity level of” (95 percent confidence)

El HLO99    Elemental t-test “homogenity level of” (99 percent confidence)

 

El HIRO60   Elemental t-test “homogenity ideal range of” (60 percent confidence)

El HIRO80   Elemental t-test “homogenity ideal range of” (80 percent confidence)

El HIRO90   Elemental t-test “homogenity ideal range of” (90 percent confidence)

El HIRO95   Elemental t-test “homogenity ideal range of” (95 percent confidence)

El HIRO99   Elemental t-test “homogenity ideal range of” (99 percent confidence)

 

El HARO60   Elemental t-test “homogenity actual range of” (60 percent confidence)

El HARO80   Elemental t-test “homogenity actual range of” (80 percent confidence)

El HARO90   Elemental t-test “homogenity actual range of” (90 percent confidence)

El HARO95   Elemental t-test “homogenity actual range of” (95 percent confidence)

El HARO99   Elemental t-test “homogenity actual range of” (99 percent confidence)

 

El CDL60    Elemental t-test detection limit (60 percent confidence)

El CDL80    Elemental t-test detection limit (80 percent confidence)

El CDL90    Elemental t-test detection limit (90 percent confidence)

El CDL95    Elemental t-test detection limit (95 percent confidence)

El CDL99    Elemental t-test detection limit (99 percent confidence)

 

El CAS60    Elemental t-test analytical sensitivity (60% CI for major elements)

El CAS80    Elemental t-test analytical sensitivity (80% CI for major elements)

El CAS90    Elemental t-test analytical sensitivity (90% CI for major elements)

El CAS95    Elemental t-test analytical sensitivity (95% CI for major elements)

El CAS99    Elemental t-test analytical sensitivity (99% CI for major elements)

 

See the section above entitled: Calculation Options | Calculate Detection Limits and Analytical Sensitivity for a detailed description of the homogeneity calculation output.