Coupling
Specifies how the digitizer couples the input signal for the selected channel(s).
Note A certain amount of delay is required for the coupling capacitor to charge after changing vertical coupling from DC to AC. This delay is typically as follows:
Low Impedance Source—150 ms
10X Probe—1.5 s
100X Probe—15 s