[SENSe:]FREQuency:VOLTage:RANGe

From Keysight 34970A/34972A

[SENSe:]FREQuency:VOLTage:RANGe
[SENSe:]PERiod:VOLTage:RANGe

Syntax

[SENSe:]FREQuency:VOLTage:RANGe {<range>|MIN|MAX}[,(@<ch_list>)]
[SENSe:]FREQuency:VOLTage:RANGe? [{(@<ch_list>)|MIN|MAX}]

[SENSe:]PERiod:VOLTage:RANGe {<range>|MIN|MAX}[,(@<ch_list>)]
[SENSe:]PERiod:VOLTage:RANGe? [{(@<ch_list>)|MIN|MAX}]

Description

These commands select the voltage range for frequency and period measurements on the specified channels.

Used With:

  • 34901A   20 Channel Multiplexer (2/4-wire) Module
  • 34902A   16 Channel Multiplexer (2/4-wire) Module
  • 34908A   40 Channel Single-Ended Multiplexer Module

Parameters

Name Type Range of Values Default Value
<range> Discrete

Desired range in volts:

100 mV (MIN)
1 V
10 V
100 V
1000 V (MAX)

This is a required parameter.
<ch_list> Channel List

One or more channels, as shown:

(@310) - channel 10 on the module in slot 300.

(@305:310) - channels 05 through 10 on the module in slot 300.

(@202:207,209,302:308) - channels 02 through 07 and 09 on the module in slot 200 and channels 02 through 08 on the module in slot 300.

If you omit the optional <ch_list> parameter, this command applies to the currently defined scan list.

Remarks

  • Selecting a discrete range will disable autoranging on the specified channels.
  • If the input signal is greater than can be measured on the selected range (manual ranging), the instrument gives an overload indication: "±OVLD" from the front panel or "±9.9E+37" from the remote interface.
  • The CONFigure and MEASure? commands automatically enable autoranging if the first parameter is AUTO, DEF or omitted.
  • Because frequency and period are related functions, changing a measurement parameter for one function will also change the corresponding parameter for the other function.
  • The instrument enables autoranging after a Factory Reset (*RST command). An Instrument Preset (SYSTem:PRESet command) or Card Reset (SYSTem:CPON command) does not change the setting.

Return Format

The query returns the voltage range in the form "+1.00000000E+01".

Examples

The following command selects the 10 volt range for frequency measurements.

FREQ:VOLT:RANG 10

 

The following query returns the voltage range selected.

FREQ:VOLT:RANG?

Typical Response: +1.00000000E+01

See Also

SENSe Subsystem Introduction

CONFigure:FREQuency

CONFigure:PERiod

[SENSe:]FREQuency:VOLTage:RANGe:AUTO

[SENSe:]PERiod:VOLTage:RANGe:AUTO