[SENSe:]FREQuency:VOLTage:RANGe
[SENSe:]FREQuency:VOLTage:RANGe
[SENSe:]PERiod:VOLTage:RANGe
Syntax
[SENSe:]FREQuency:VOLTage:RANGe {<range>|MIN|MAX}[,(@<ch_list>)]
[SENSe:]FREQuency:VOLTage:RANGe? [{(@<ch_list>)|MIN|MAX}]
[SENSe:]PERiod:VOLTage:RANGe {<range>|MIN|MAX}[,(@<ch_list>)]
[SENSe:]PERiod:VOLTage:RANGe? [{(@<ch_list>)|MIN|MAX}]
Description
These commands select the voltage range for frequency and period measurements on the specified channels.
Used With:
- 34901A 20 Channel Multiplexer (2/4-wire) Module
- 34902A 16 Channel Multiplexer (2/4-wire) Module
- 34908A 40 Channel Single-Ended Multiplexer Module
Parameters
Name Type Range of Values Default Value <range> Discrete Desired range in volts:
100 mV (MIN)
1 V
10 V
100 V
1000 V (MAX)This is a required parameter. <ch_list> Channel List One or more channels, as shown:
(@310) - channel 10 on the module in slot 300.
(@305:310) - channels 05 through 10 on the module in slot 300.
(@202:207,209,302:308) - channels 02 through 07 and 09 on the module in slot 200 and channels 02 through 08 on the module in slot 300.
If you omit the optional <ch_list> parameter, this command applies to the currently defined scan list.
Remarks
- Selecting a discrete range will disable autoranging on the specified channels.
- If the input signal is greater than can be measured on the selected range (manual ranging), the instrument gives an overload indication: "±OVLD" from the front panel or "±9.9E+37" from the remote interface.
- The CONFigure and MEASure? commands automatically enable autoranging if the first parameter is AUTO, DEF or omitted.
- Because frequency and period are related functions, changing a measurement parameter for one function will also change the corresponding parameter for the other function.
- The instrument enables autoranging after a Factory Reset (*RST command). An Instrument Preset (SYSTem:PRESet command) or Card Reset (SYSTem:CPON command) does not change the setting.
Return Format
The query returns the voltage range in the form "+1.00000000E+01".
Examples
The following command selects the 10 volt range for frequency measurements.
FREQ:VOLT:RANG 10
The following query returns the voltage range selected.
FREQ:VOLT:RANG?
Typical Response: +1.00000000E+01