Factory Reset State
Factory Reset State
The following tables show the state of the instrument after a FACTORY RESET from the Sto/Rcl menu, or *RST from the remote interface.
|
Measurement Configuration |
Factory Reset State |
|
Function |
DC Volts |
|
Range |
Autorange |
|
Resolution |
5½ Digits |
|
Integration Time |
1 PLC |
|
Input Resistance |
10 MΩ (fixed for all DCV ranges) |
|
Channel Delay |
Automatic Delay |
|
Totalizer Reset Mode |
Count Not Reset When Read |
|
Totalizer Edge Detect |
Rising Edge |
|
Scanning Operations |
Factory Reset State |
|
Scan List |
Empty |
|
Reading Memory |
All Readings are Cleared |
|
Min, Max, and Average |
Not Changed |
|
Scan Trigger Source |
Immediate |
|
Scan Interval (used with TRIGger:SOURce TIMer) |
10 Seconds |
|
Scan Count |
1 |
|
Scan Reading Format |
Reading Only (No Units, Channel, Time) |
|
Monitor in Progress |
Stopped |
|
Mx+B Scaling |
Factory Reset State |
|
Gain Factor ("M") |
1 |
|
Offset Factor ("B") |
0 |
|
Scale Label |
VDC |
|
Alarm Limits |
Factory Reset State |
|
Alarm Queue |
Not Cleared |
|
Alarm State |
Off |
|
HI and LO Alarm Limits |
0 |
|
Alarm Output |
Alarm 1 |
|
Alarm Output Configuration |
Latched Mode |
|
Alarm Output State |
Output Lines are Cleared |
|
Alarm Output Slope |
Fail = Low |
|
Module Hardware |
Factory Reset State |
|
34901A, 34902A, 34908A |
All Channels Open |
|
34903A, 34904A |
All Channels Open |
|
34905A, 34906A |
Channels s11 and s21 Selected |
|
34907A |
Both DIO Ports = Input, Totalizer Count = 0, Both DACs = 0 VDC |
|
System-Related Operations |
Factory Reset State |
|
Display State |
On |
|
Error Queue |
Errors Not Cleared |
|
Stored States |
No Change |