Factory Reset State

From Keysight 34970A/34972A

Factory Reset State

The following tables show the state of the instrument after a FACTORY RESET from the Sto/Rcl menu, or *RST from the remote interface.

 

Measurement Configuration

Factory Reset State

Function

DC Volts

Range

Autorange

Resolution

5½ Digits

Integration Time

1 PLC

Input Resistance

10 MΩ (fixed for all DCV ranges)

Channel Delay

Automatic Delay

Totalizer Reset Mode

Count Not Reset When Read

Totalizer Edge Detect

Rising Edge

 

 

Scanning Operations

Factory Reset State

Scan List

Empty

Reading Memory

All Readings are Cleared

Min, Max, and Average

Not Changed

Scan Trigger Source

Immediate

Scan Interval (used with TRIGger:SOURce TIMer)

10 Seconds

Scan Count

1

Scan Reading Format

Reading Only (No Units, Channel, Time)

Monitor in Progress

Stopped

 

 

Mx+B Scaling

Factory Reset State

Gain Factor ("M")

1

Offset Factor ("B")

0

Scale Label

VDC

 

 

Alarm Limits

Factory Reset State

Alarm Queue

Not Cleared

Alarm State

Off

HI and LO Alarm Limits

0

Alarm Output

Alarm 1

Alarm Output Configuration

Latched Mode

Alarm Output State

Output Lines are Cleared

Alarm Output Slope

Fail = Low

 

 

Module Hardware

Factory Reset State

34901A, 34902A, 34908A

All Channels Open

34903A, 34904A

All Channels Open

34905A, 34906A

Channels s11 and s21 Selected

34907A

Both DIO Ports = Input, Totalizer Count = 0,  Both DACs = 0 VDC

 

 

System-Related Operations

Factory Reset State

Display State

On

Error Queue

Errors Not Cleared

Stored States

No Change