[SENSe:]TEMPerature:TRANsducer:TCouple:TYPE

From Keysight 34970A/34972A

[SENSe:]TEMPerature:TRANsducer:TCouple:TYPE

Syntax

[SENSe:]TEMPerature:TRANsducer:TCouple:TYPE <type>[,(@<ch_list>)]

[SENSe:]TEMPerature:TRANsducer:TCouple:TYPE? [(@<ch_list>)]

Description

This command selects the thermocouple type to use for measurements on the specified channels.

Used With:

  • 34901A   20 Channel Multiplexer (2/4-wire) Module
  • 34902A   16 Channel Multiplexer (2/4-wire) Module
  • 34908A   40 Channel Single-Ended Multiplexer Module

Parameters

Name Type Range of Values Default Value
<type> Discrete {B|E|J|K|N|R|S|T} This is a required parameter.
<ch_list> Channel List

One or more channels, as shown:

(@310) - channel 10 on the module in slot 300.

(@305:310) - channels 05 through 10 on the module in slot 300.

(@202:207,209,302:308) - channels 02 through 07 and 09 on the module in slot 200 and channels 02 through 08 on the module in slot 300.

If you omit the optional <ch_list> parameter, this command applies to the currently defined scan list.

Remarks

  • Thermocouple measurements require a reference junction temperature (see [SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:TYPE command). For the reference junction temperature, you can use an internal measurement on the module's terminal block (34901A only), an external thermistor or RTD measurement, or a known fixed junction temperature. By default, a fixed reference junction temperature of 0.0 °C is used.
  • The instrument sets the thermocouple type to "J" after a Factory Reset (*RST command). An Instrument Preset (SYSTem:PRESet command) or Card Reset (SYSTem:CPON command) does not change the setting.

Return Format

The query returns the thermocouple type as a quoted letter for each channel in the list. Multiple channels are separated by commas.

Examples

The following command sets the thermocouple to J for the channels shown.

TEMP:TRAN:TCouple:TYPE J,(@201:204)

 

The following query returns the thermocouple type for the channels shown.

TEMP:TRAN:TC:TYPE? (@101,103,202:204,301)

Typical Response: B,R,J,J,J,K

See Also

SENSe Subsystem Introduction