CALCulate:SCALe:OFFSet:NULL
CALCulate:SCALe:OFFSet:NULL
Syntax
CALCulate:SCALe:OFFSet:NULL [(@<ch_list>)]
Description
This command makes an immediate null measurement on the specified channels and stores it as the offset ("B") for subsequent measurements. This allows you to adjust for voltage or resistive offsets through your wiring to the point of the measurement.
Parameters
Name Type Range of Values Default Value <ch_list> Channel List One or more channels, as shown:
(@310) - channel 10 on the module in slot 300.
(@305:310) - channels 05 through 10 on the module in slot 300.
(@202:207,209,302:308) - channels 02 through 07 and 09 on the module in slot 200 and channels 02 through 08 on the module in slot 300.
If you omit the optional <ch_list> parameter, this command applies to the currently defined scan list.
Example
The following command makes an immediate null measurement on channels 206 through 210.
CALC:SCAL:OFFS:NULL (@206:210)