Edit TH-03 Channel

PicoLog

Edit TH-03 Channel

Top  Previous  Next

Accessed via the Edit button in the TH03 Channels dialog box.

th03edms

This dialog box is used to enter the details for a channel.

Name.  This is the name of the channel. This name will appear on all reports.

Sensor type.  This is the sensor connected to the channel. There are seven choices:

blueballEL015 (low temp) is a thermistor sensor suitable for measuring temperatures in the range -30 to 70°C.
blueballEL029 door switch is a magnetic sensor, which senses if a door is open or closed.
blueballEL031 light is a sensor suitable for measuring light intensity.
blueballEL039 (low temp) is a thermistor sensor suitable for measuring temperatures in the range -40 to 30°C.
blueballMedium temp is a thermistor sensor suitable for measuring temperatures in the range 0 to 70°C.
blueballHigh temp is a thermistor sensor suitable for measuring temperatures in the range 0 to 300°C.
blueballUnicam is reserved for special uses

 

Filter enable.  If you want to use a filter to reduce the effects of electrical noise, check this box.

Filter factor.  Each time a reading is taken, the filtered value is updated using the equation:

new value = old value + (reading - old value) / (filter factor)

Filtered values are more stable, but respond more slowly to real changes in temperature. The higher the factor, the stronger the filtering effect will be.

Options.  To set the options for a parameter, click the Options... button.