Extra Camera Settings

WinFluor V4.0.3

Recording > Images + Signals > Extra Camera Settings

 

The special features of certain cameras can be adjusted before beginning recording by selecting

 

Setup > Extra Camera Settings

 

to open the Camera Settings dialog box.

 

 

Note that these settings only take effect where the feature is supported by the camera.

 

Cooling (Andor cameras only)

 

CCD cooling can be enabled/disabled by ticking/unticking the Peltier Cooling On option.

 

Camera Fan (Andor & Hamamatsu cameras)

 

The camera air cooling fan can be stopped (or slowed) to reduce vibration by changing the Fan Mode option to Off or Low.  Note that if the fan is turned off for sustained periods of time without a substitute cooling system such as water cooling being available the camera may overheat and malfunction.

 

Readout A/D Gain

 

The pre-amplifier gain of the CCD readout analogue/digital converter can be selected from the Readout A/D Gain list.

 

Vertical Line Shift Time (Andor Ixon cameras)

 

The clock period driving the rate at which lines of pixels are shifted vertically down the CCD sensor towards the readout line can be selected from the Vertical Line Shift Time list.

 

LightSpeed Mode (Photometrics Evolve 512 Delta)

 

Enables/disables LightSpeed CCD readout mode of the Photometrics Evolve 512 Delta camera, allowing faster CCD readout rates from CCD sub-regions (Note a physical CCD mas is required to define the sub-region).