NI PXI-4130
The NI PXI-4130 consists of two output channels. Channel 0, also referred to as the utility channel, is a single-quadrant power supply. Channel 1, the SMU channel, is a four-quadrant, bipolar power source-measure unit. The following table lists the DC voltage ranges supported by each channel.
Channel | Range | |
---|---|---|
Output | Measurement | |
0 | +6 V | +6 V |
1 | ±20 V | ±20 V |
±6 V | ||
These channels support 5% overranging when overranging is enabled. |
The following table lists DC current ranges supported by the NI PXI-4130.
Channel | Range | |
---|---|---|
Output | Measurement | |
0 | 1 A | 1 A |
1 | 200 μA | 200 μA |
2 mA | 2 mA | |
20 mA | 20 mA | |
200 mA* | 200 mA* | |
2 A* | 2 A* | |
*See Internal and Auxiliary Power for limitations when this device is under internal power. | ||
These channels support 5% overranging when overranging is enabled. |
The following diagrams illustrate the output voltage and current capabilities of each channel on the NI PXI-4130:
For more information about the NI PXI-4130 specifications, refer to Related Documentation.