NI PXI-4130

NI DC Power Supply & SMU

NI PXI-4130

The NI PXI-4130 consists of two output channels. Channel 0, also referred to as the utility channel, is a single-quadrant power supply. Channel 1, the SMU channel, is a four-quadrant, bipolar power source-measure unit. The following table lists the DC voltage ranges supported by each channel.

Channel Range
OutputMeasurement
0+6 V+6 V
1±20 V±20 V
±6 V
These channels support 5% overranging when overranging is enabled.

The following table lists DC current ranges supported by the NI PXI-4130.

Channel Range
OutputMeasurement
01 A1 A
1200 μA200 μA
2 mA2 mA
20 mA20 mA
200 mA*200 mA*
2 A*2 A*
*See Internal and Auxiliary Power for limitations when this device is under internal power.
These channels support 5% overranging when overranging is enabled.

The following diagrams illustrate the output voltage and current capabilities of each channel on the NI PXI-4130:

   

For more information about the NI PXI-4130 specifications, refer to Related Documentation.