ABORt

From Keysight 34970A/34972A

ABORt

Syntax

ABORt

Description

This command aborts a measurement in progress from a scan.

Remarks

  • If a scan is in progress when the command is received, the scan will not be completed and you cannot resume the scan from where it left off. Note that if you initiate a new scan, all readings are cleared from memory.
  • The *RST command will abort a measurement, clear the scan list, and set all measurement parameters to their factory settings. The Instrument Preset (SYSTem:PRESet command) also aborts a measurement but it does not clear the scan list.

Example

The following command aborts the measurement in progress.

ABOR

See Also

*RST

SYSTem:PRESet