[SENSe:]TEMPerature:RJUNction?
[SENSe:]TEMPerature:RJUNction?
Syntax
[SENSe:]TEMPerature:RJUNction? [(@<ch_list>)]
Description
This query returns he internal reference junction temperature on the specified channels. This is useful only for an internal reference source).
Parameters
Name Type Range of Values Default Value <ch_list> Channel List One or more channels, as shown:
(@310) - channel 10 on the module in slot 300.
(@305:310) - channels 05 through 10 on the module in slot 300.
(@202:207,209,302:308) - channels 02 through 07 and 09 on the module in slot 200 and channels 02 through 08 on the module in slot 300.
If you omit the optional <ch_list> parameter, this command applies to the currently defined scan list.
Remarks
- This query returns the reference temperature in degrees Celsius, regardless of the temperature units currently selected (see UNIT:TEMPerature command).
Return Format
The query returns a number in the form +2.89753100E+01.
Example
The following query returns the temperature of the internal reference junction on channels 03, 04, and 05 in slot 200.
TEMP:RJUN? (@203:205) !Always returns result in degrees Celsius
Typical Response: +2.35212231E+01,+2.37701293E+01,+2.38291321E+01