DIAGnostic:RELay:CYCLes:CLEar

From Keysight 34970A/34972A

DIAGnostic:RELay:CYCLes:CLEar

Syntax

DIAGnostic:RELay:CYCLes:CLEar (@<ch_list>)

Description

This command resets the cycle count on the specified channels.

Parameters

Name Type Range of Values Default Value
<ch_list> Channel List

One or more channels, as shown:

(@310) - channel 10 on the module in slot 300.

(@305:310) - channels 05 through 10 on the module in slot 300.

(@202:207,209,302:308) - channels 02 through 07 and 09 on the module in slot 200 and channels 02 through 08 on the module in slot 300.

If this parameter is omitted, the command applies to the current scan list.

Remarks

  • The instrument must be unsecured in order to reset the cycle count. See the CALibration:SECure:STATe command for more information on unsecuring the instrument.
  • On the RF Multiplexer modules (34905A, 34906A), each bank consists of two leaf relays and one tree relay (see DIAGnostic:RELay:CYCLes? command). Clearing the cycle count on a specific channel will clear the count on all three relays in the corresponding bank.

Example

The following command clears the cycle count on channels 03 and 13 in slot 100.

DIAG:REL:CYCL:CLE (@103,113)

See Also

DIAGnostic Subsystem Introduction

CALibration:SECure:STATe

DIAGnostic:RELay:CYCLes?